SYRINGE PUMP SPECIFICATIONS
Infusion rate range |
10 ml syringe ------ 0.1-300 ml/h
|
50/60 ml syringe ------ 0.1-1200 ml/h
|
20/30 ml syringe ------ 0.1-500 ml/h |
100 ml syringe ------ 0.1-1500 ml/h |
Bolus / Purge rate |
10 ml syringe ------ 240 ml/h
|
50/60 ml syringe ------ 960 ml/h
|
20/30 ml syringe ------ 400 ml/h |
100 ml syringe ------ 1200 ml/h |
Bolus volume |
Programmable from 1 to 10 ml |
low rate increments |
0.1-99 ml/h at 0.1 ml/h increments
|
100-1500 ml/h at 1ml/h increments |
Accuracy |
± 5% with syringe |
Occlusion pressure |
0.04 Mpa – 0.12 Mpa / 5.8 PSI – 17.4 PSI |
Alarm volume |
High, Medium, Low |
Safety protection |
Automatic keypad lock
|
Audible & visual alarms |
Alarms |
NEAR EMPTY (syringe almost empty)
|
Syringe disengaged |
COMPLETE (Infusion completed)
|
Mechanical error |
OCCLUSION
|
Power source change |
LOW BATT |
|
Power supply |
AC 100-240V 50/60 Hz; DC 12V |
Battery |
Built-in rechargeable sealed Ni-MH
|
Pump will operate 3 hours min. at 5 ml/h with fully charged battery |
Net weight |
1.85 Kg / 4 lbs |
Dimensions |
335 (L) x 142 (W) x 120 (H) mm / 13” (L) x 51½, ” (W) x 3 ½“ (H) |